Autoclave
Hanshin Autoclave
Pre-vacuum Air Removal & Post-vacuum Drying
Hanshin autoclave is an economic B-class sterilizer that is designed to remove air in the chamber through the pre-vacuum process. Steam penetration ensures proper sterilization of all instruments. Hanshin autoclave has built-in exterior small steam generator which is newly-developed by Hanshin Medical
Type

HS-2522BL

HS-3041BL
Overall Dimension
HS-2522BL
W513 x D625 x H442 mm
HS-3041BL
W615 x D757 x H505 mm
Chamber
HS-2522BL
ID Ø 255 x L447 mm, Vol: 22ℓ, Material: STS 316L
HS-3041BL
W302 x D450 x H302 mm, Vol: 40.5ℓ, Material: STS 316L
Max. Working pressure: 2.4 bar / Min. Working Pressure: -1.00 bar
Max. Working Temperature: 138.0 °C / Test Pressure: 5.13 bar
Tray
HS-2522BL
W195 x D375 x H18 mm, Qty. 4ea
HS-3041BL
W288 x D410 x H18 mm, Qty. 4ea
Operating Temperature
121.0°C ~ 135.0°C (display resolution : 0.1 °C)
Reservoir/Collection
HS-2522BL
Vol: 6.0ℓ / 6.0ℓ
HS-3041BL
W302 x D450 x H302 mm, Vol: 40.5ℓ, Material: STS 316L
Reservoir Minimum Water Vol : 2.7ℓ
Type

HS-3041BL
How many years of experience do you have?
What services do you offer?
Are you licensed and Insured?
What is the construction process?
What happens if the project goes over budget?
How long does it take to get an estimate?
Will I be able to request changes during the project?
BLseries

1
Steam Generator
This device is installed in the external side and injects the saturated steam as a sterilant into the chamber generated instantly with the feed water supplied from the reservoir.

2
Chamber
By adopting an integral chamber unleaded, the durability is excellent and a treble such as leaking of pressurised vapor and vacuum during progressing a cycle is not occurred accidently.

3
Reservoir and Collection Container
The feed water for generating the saturated steam as a sterilant in the reservoir and the used water(condensate) after completion of a cycle is stored until draining.

4
Safety Valve
This valve is opened automatically and releases the overpressure(more that 2.8 bar) safely formed in the chamber due to an abnormal operation of steriliser and is set to 2.8 bar pressure(Capacity; 129 kg/hr, Minimum operating pressure margin; 10%) at the maximum temperature of 150°C.
This is the finished product imported form USA, which received approvals of ASME Section VIII Div 1 Pressure Vessels in America and Pressure Equipment Directive 97/23/EC(PED), Directive 2002/95/EC(RoHS) in Europe.

5
Control Panel
The parameters of the selected cycle program and the state of cycle progressing are displayed digitally in real time on this LCD monitor to check the control the operation of steriliser.
The operator can operate the steriliser and change a cycle program by using the functional buttons arranged on this control panel. The water level in the reservoir and collection container and overheat state are indicated visually by the LEDs on it.

6
Vacuum Pump
This pump removes air in the chamber enough to the set vacuum level and allow the loads to be stressed the dried completely under the optimised condition.

7
Panel Printer
The temperature/pressure graph and the cycle performance record through a cycle progress are printed out.

8
Manual Reset Switch of Thermostat
The thermostat cuts off electric power to the heater automatically when the heater is overheated higher than the set temperature, the operators should press this reset switch manually to recover power supplying to the heater after removing the cause and cooling the heater.

9
USB & LAN Port
The user can download or backup the cycle data from the controller of steriliser in the USB flash memory and monitor the operation of terser on the external PC by connecting the USB or LAN cable to the these ports. Moreover the communication program including an ubiquitous function provided allows user to manage the operation of plural(maximum 15 sets) sterilisers at once. (These functions are provided as an optional specification)
10
Air Filter (HEPA)
The bacteria and foreign substances included in the air flowed into the chamber are filtered with a high efficiency for preventing the stressed loads from recontamination in the chamber.
11
Drain Valve of Reservoir and Collection Container
When requiring the water in the reservoir to drain and the water in the collection container is in full level, the operator open these valves for draining the water.
Steam Generator

This device is installed in the external side and injects the saturated steam as a sterilant into the chamber generated instantly with the feed water supplied from the reservoir.
Chamber

By adopting an integral chamber unleaded, the durability is excellent and a treble such as leaking of pressurised vapor and vacuum during progressing a cycle is not occurred accidently.
Reservoir and Collection Container

The feed water for generating the saturated steam as a sterilant in the reservoir and the used water(condensate) after completion of a cycle is stored until draining.
Safety Valve

This valve is opened automatically and releases the overpressure(more that 2.8 bar) safely formed in the chamber due to an abnormal operation of steriliser and is set to 2.8 bar pressure(Capacity; 129 kg/hr, Minimum operating pressure margin; 10%) at the maximum temperature of 150°C.
This is the finished product imported form USA, which received approvals of ASME Section VIII Div 1 Pressure Vessels in America and Pressure Equipment Directive 97/23/EC(PED), Directive 2002/95/EC(RoHS) in Europe.
Control Panel

The parameters of the selected cycle program and the state of cycle progressing are displayed digitally in real time on this LCD monitor to check the control the operation of steriliser.
The operator can operate the steriliser and change a cycle program by using the functional buttons arranged on this control panel. The water level in the reservoir and collection container and overheat state are indicated visually by the LEDs on it.
Vacuum Pump

This pump removes air in the chamber enough to the set vacuum level and allow the loads to be stressed the dried completely under the optimised condition.
Panel Printer

The temperature/pressure graph and the cycle performance record through a cycle progress are printed out.
Manual Reset Switch of Thermostat

The thermostat cuts off electric power to the heater automatically when the heater is overheated higher than the set temperature, the operators should press this reset switch manually to recover power supplying to the heater after removing the cause and cooling the heater.
USB & LAN Port
The user can download or backup the cycle data from the controller of steriliser in the USB flash memory and monitor the operation of terser on the external PC by connecting the USB or LAN cable to the these ports. Moreover the communication program including an ubiquitous function provided allows user to manage the operation of plural(maximum 15 sets) sterilisers at once. (These functions are provided as an optional specification)
Air Filter (HEPA)
The bacteria and foreign substances included in the air flowed into the chamber are filtered with a high efficiency for preventing the stressed loads from recontamination in the chamber.
Drain Valve of Reservoir and Collection Container
When requiring the water in the reservoir to drain and the water in the collection container is in full level, the operator open these valves for draining the water.
Hanshin Autoclave B-Class
Futhermore the sterilizing cycle programs have a unique drying process that the load is dried in a vacuum after sterilizing process, and the dryness for loads passed through this drying process is complied with the strict requirement of the international standards.
The user can download or backup those data in the USB memory and monitor the operation of sterilizer on the external PC by connecting the USB or LAN cable to the built-in communication ports. Moreover the communication program including an ubiquitous function provided allows user to manage the operation of plural(maximum 15 sets) sterilizers at once. (optional specification)
Except those fundamental safety functions, these sterilizers adopt a full proof safety system with the safety valve, protection program for over temperature and pressure, circuit breaker etc. and therefore the safety for operators and working area will be guaranteed certainly during operating.
High Speed Mixing
Full Mouth Tray Injection in 10seconds
· More working time with a fast dispense.
· Completion of injecting to the tray in 10seconds at the touch of a button.
Dispense Speed Value


Hardening Time Alarming Function
30sec Interval, 8minutes Max
· Set an alarm to the hardening time of the impression material used (8minutes max)
· Convenient dial button
· Hardening time settable by users according to the external environment (temperature and hardener)
Alarming Function
Alarm set > Dispense alarm rings > Remove the impression body