Autoclave

Hanshin Autoclave

Pre-vacuum Air Removal & Post-vacuum Drying

Hanshin autoclave is an economic B-class sterilizer that is designed to remove air in the chamber through the pre-vacuum process. Steam penetration ensures proper sterilization of all instruments. Hanshin autoclave has built-in exterior small steam generator which is newly-developed by Hanshin Medical

Type

HS-2522BL

HS-3041BL

Overall Dimension

HS-2522BL

W513 x D625 x H442 mm

HS-3041BL

W615 x D757 x H505 mm

Chamber

HS-2522BL

ID Ø 255 x L447 mm, Vol: 22ℓ, Material: STS 316L

HS-3041BL

W302 x D450 x H302 mm, Vol: 40.5ℓ, Material: STS 316L

Max. Working pressure: 2.4 bar / Min. Working Pressure: -1.00 bar
Max. Working Temperature: 138.0 °C / Test Pressure: 5.13 bar

Tray

HS-2522BL

W195 x D375 x H18 mm, Qty. 4ea

HS-3041BL

W288 x D410 x H18 mm, Qty. 4ea

Operating Temperature

121.0°C ~ 135.0°C (display resolution : 0.1 °C)

Reservoir/Collection

HS-2522BL

Vol: 6.0ℓ / 6.0ℓ

HS-3041BL

W302 x D450 x H302 mm, Vol: 40.5ℓ, Material: STS 316L

Reservoir Minimum Water Vol : 2.7ℓ

HS-2522BL
HS-3041BL
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.

BLseries

1

Steam Generator

This device is installed in the external side and injects the saturated steam as a sterilant into the chamber generated instantly with the feed water supplied from the reservoir.

2

Chamber

By adopting an integral chamber unleaded, the durability is excellent and a treble such as leaking of pressurised vapor and vacuum during progressing a cycle is not occurred accidently.

3

Reservoir and Collection Container

The feed water for generating the saturated steam as a sterilant in the reservoir and the used water(condensate) after completion of a cycle is stored until draining.

4

Safety Valve

This valve is opened automatically and releases the overpressure(more that 2.8 bar) safely formed in the chamber due to an abnormal operation of steriliser and is set to 2.8 bar pressure(Capacity; 129 kg/hr, Minimum operating pressure margin; 10%) at the maximum temperature of 150°C.

This is the finished product imported form USA, which received approvals of ASME Section VIII Div 1 Pressure Vessels in America and Pressure Equipment Directive 97/23/EC(PED), Directive 2002/95/EC(RoHS) in Europe.

5

Control Panel

The parameters of the selected cycle program and the state of cycle progressing are displayed digitally in real time on this LCD monitor to check the control the operation of steriliser.

The operator can operate the steriliser and change a cycle program by using the functional buttons arranged on this control panel. The water level in the reservoir and collection container and overheat state are indicated visually by the LEDs on it.

6

Vacuum Pump

This pump removes air in the chamber enough to the set vacuum level and allow the loads to be stressed the dried completely under the optimised condition.

7

Panel Printer

The temperature/pressure graph and the cycle performance record through a cycle progress are printed out.

8

Manual Reset Switch of Thermostat

The thermostat cuts off electric power to the heater automatically when the heater is overheated higher than the set temperature, the operators should press this reset switch manually to recover power supplying to the heater after removing the cause and cooling the heater.

9

USB & LAN Port

The user can download or backup the cycle data from the controller of steriliser in the USB flash memory and monitor the operation of terser on the external PC by connecting the USB or LAN cable to the these ports. Moreover the communication program including an ubiquitous function provided allows user to manage the operation of plural(maximum 15 sets) sterilisers at once. (These functions are provided as an optional specification)

10

Air Filter (HEPA)

The bacteria and foreign substances included in the air flowed into the chamber are filtered with a high efficiency for preventing the stressed loads from recontamination in the chamber.

11

Drain Valve of Reservoir and Collection Container

When requiring the water in the reservoir to drain and the water in the collection container is in full level, the operator open these valves for draining the water.

1

Steam Generator

This device is installed in the external side and injects the saturated steam as a sterilant into the chamber generated instantly with the feed water supplied from the reservoir.

2

Chamber

By adopting an integral chamber unleaded, the durability is excellent and a treble such as leaking of pressurised vapor and vacuum during progressing a cycle is not occurred accidently.

3

Reservoir and Collection Container

The feed water for generating the saturated steam as a sterilant in the reservoir and the used water(condensate) after completion of a cycle is stored until draining.

4

Safety Valve

This valve is opened automatically and releases the overpressure(more that 2.8 bar) safely formed in the chamber due to an abnormal operation of steriliser and is set to 2.8 bar pressure(Capacity; 129 kg/hr, Minimum operating pressure margin; 10%) at the maximum temperature of 150°C.

This is the finished product imported form USA, which received approvals of ASME Section VIII Div 1 Pressure Vessels in America and Pressure Equipment Directive 97/23/EC(PED), Directive 2002/95/EC(RoHS) in Europe.

5

Control Panel

The parameters of the selected cycle program and the state of cycle progressing are displayed digitally in real time on this LCD monitor to check the control the operation of steriliser.

The operator can operate the steriliser and change a cycle program by using the functional buttons arranged on this control panel. The water level in the reservoir and collection container and overheat state are indicated visually by the LEDs on it.

6

Vacuum Pump

This pump removes air in the chamber enough to the set vacuum level and allow the loads to be stressed the dried completely under the optimised condition.

7

Panel Printer

The temperature/pressure graph and the cycle performance record through a cycle progress are printed out.

8

Manual Reset Switch of Thermostat

The thermostat cuts off electric power to the heater automatically when the heater is overheated higher than the set temperature, the operators should press this reset switch manually to recover power supplying to the heater after removing the cause and cooling the heater.

9

USB & LAN Port

The user can download or backup the cycle data from the controller of steriliser in the USB flash memory and monitor the operation of terser on the external PC by connecting the USB or LAN cable to the these ports. Moreover the communication program including an ubiquitous function provided allows user to manage the operation of plural(maximum 15 sets) sterilisers at once. (These functions are provided as an optional specification)

10

Air Filter (HEPA)

The bacteria and foreign substances included in the air flowed into the chamber are filtered with a high efficiency for preventing the stressed loads from recontamination in the chamber.

11

Drain Valve of Reservoir and Collection Container

When requiring the water in the reservoir to drain and the water in the collection container is in full level, the operator open these valves for draining the water.

Hanshin Autoclave B-Class

This practical B-Class sterilizer is designed to reduce cost remaining B-Class performance, therefore all the medical devices and supplies within the sterilizable articles of B-Class can be sterilized economically, and their rectangular chamber allows user to use wide usable space in the chamber.
This sterilizer is an economic B-class sterilizer that is designed to remove air in the chamber and loads through the pre-vacuum process for which all the medical instruments and devices able to be penetrated steam can be sterilized effectively.

Futhermore the sterilizing cycle programs have a unique drying process that the load is dried in a vacuum after sterilizing process, and the dryness for loads passed through this drying process is complied with the strict requirement of the international standards.
The used water after completion of a sterilizing cycle is collected in the collecting container designed separately with the reservoir, therefore the saturated steam generated by a clean water always can be supplied to perform every sterilizing cycle.
By adopting a integral chamber unwelded, the durability is excellent and a trouble such as leaking of pressurized vapor and vacuum during progressing a cycle is not occurred accidently.
The cycle performance records executed at least for 11 years are stored in the control unit in order. The user can automatically search the results of sterilization cycle performed previously by input the date performed cycle or the cycle count number and prints out the result record by the built-in panel printer.

The user can download or backup those data in the USB memory and monitor the operation of sterilizer on the external PC by connecting the USB or LAN cable to the built-in communication ports. Moreover the communication program including an ubiquitous function provided allows user to manage the operation of plural(maximum 15 sets) sterilizers at once. (optional specification)
The cycle is not started unless the door has been locked by the interlock system, the door is not opened under no electric power and also there exists more pressure than the set level in the chamber.

Except those fundamental safety functions, these sterilizers adopt a full proof safety system with the safety valve, protection program for over temperature and pressure, circuit breaker etc. and therefore the safety for operators and working area will be guaranteed certainly during operating.
The optional panel printer prints out the cycle performance record for analysing of sterilization results and securing the traceability afterwards. (The panel printer is provided as an optional item)

High Speed Mixing

Full Mouth Tray Injection in 10seconds

· More working time with a fast dispense.
· Completion of injecting to the tray in 10seconds at the touch of a button.

Dispense Speed Value

Hardening Time Alarming Function

30sec Interval, 8minutes Max

· Set an alarm to the hardening time of the impression material used (8minutes max)
· Convenient dial button
· Hardening time settable by users according to the external environment (temperature and hardener)

Alarming Function

Alarm set > Dispense alarm rings > Remove the impression body

This practical B-Class sterilizer is designed to reduce cost remaining B-Class performance, therefore all the medical devices and supplies within the sterilizable articles of B-Class can be sterilized economically, and their rectangular chamber allows user to use wide usable space in the chamber.

Unique B Class small steam sterilizer Both economical & practical

This sterilizer is an economic B-class sterilizer that is designed to remove air in the chamber and loads through the pre-vacuum process for which all the medical instruments and devices able to be penetrated steam can be sterilized effectively.

Futhermore the sterilizing cycle programs have a unique drying process that the load is dried in a vacuum after sterilizing process, and the dryness for loads passed through this drying process is complied with the strict requirement of the international standards.

Unique exterior small steam generator

The built-in exterior small steam generator which is newly-developed by our advanced technology can vaporize feed water supplied from the reservoir instantly and the generated steam is injected into the chamber as a sterilizing media, therefore the chamber is less contaminated than the internal steam generating type and the usable loading space is widened because there is no heater assembly in the chamber.

Use of clean steam every cycle

The used water after completion of a sterilizing cycle is collected in the collecting container designed separately with the reservoir, therefore the saturated steam generated by a clean water always can be supplied to perform every sterilizing cycle.

Safe and durable unwelded chamber

By adopting a integral chamber unwelded, the durability is excellent and a trouble such as leaking of pressurized vapor and vacuum during progressing a cycle is not occurred accidently.

Cycle Execution & Data Storing Function

The cycle performance records executed at least for 11 years are stored in the control unit in order. The user can automatically search the results of sterilization cycle performed previously by input the date performed cycle or the cycle count number and prints out the result record by the built-in panel printer.

The user can download or backup those data in the USB memory and monitor the operation of sterilizer on the external PC by connecting the USB or LAN cable to the built-in communication ports. Moreover the communication program including an ubiquitous function provided allows user to manage the operation of plural(maximum 15 sets) sterilizers at once. (optional specification)

Protective safety system

The cycle is not started unless the door has been locked by the interlock system, the door is not opened under no electric power and also there exists more pressure than the set level in the chamber.

Except those fundamental safety functions, these sterilizers adopt a full proof safety system with the safety valve, protection program for over temperature and pressure, circuit breaker etc. and therefore the safety for operators and working area will be guaranteed certainly during operating.

Printing of the cycle performance record (optional)

The optional panel printer prints out the cycle performance record for analysing of sterilization results and securing the traceability afterwards. (The panel printer is provided as an optional item)

Overall Dimension

W615 x D757 x H505 mm

Chamber

W302 x D450 x H302 mm, Vol: 40.5, Material: STS 316L

Reservoir/Collection​

Max. Working pressure: 2.4 bar / Min. Working Pressure: -1.00 bar Max. Working Temperature: 138.0 °C / Test Pressure: 5.13 bar​

Container

Capacity; 7.5 / 7.5 (Full water level)

Tray

W288 x D410 x H18 mm, 4 ea

Operating Temperature

121.0°C ~ 135.0°C (Display Resolution; 0.1°C)

Sterilising Agent

Saturated vapor

Air Removal

Vacuum

Air Removal Device

Vacuum Pump (Diaphragm type)

Controls

Microprocessor

Display

Graphic LCD (LED backlighting)

Panel Printer (Optional)

Direct thermal printing, Printing Paper Roll: W58 x Ø40 mm, L=13 m

Power Source

AC 230V, 50/60Hz

Power Consumption

3,300 W

Net Weight

78 kg

Environment Conditions (IEC 61010-1)

* Altitudes up to 2,000 m
* Temperature range of + 5°C to + 40°C.
* Maximum Relative Humidity 80% for temperatures up to 31°Cdecreasing Iinearly to 50% relative humidity at 40°C.
* Mains supply voltage fluctuations up to ±10% of the nominal.
* Installation Category (Overvoltage Category): , Pollution Degree : 2.